SOP FOR GENERAL FUNCTION OF MICROBIOLOGICAL LAB IN PHARMACEUTICAL INDUSTRY
![]() |
| MICROBIOLOGICAL LAB |
1.0. Purpose
This SOP is required to get a general concept about the functions performed in Microbiology lab.
2.0. Scope
This procedure is applicable to the Microbiology Lab of
(Pvt.) Limited.
|
Action |
Responsibility |
|
To follow this procedure |
Microbiologist |
|
To implement this procedure |
Incharge Microbiology Lab |
Following are the main
functions that perform in Microbiology lab to get the general concept of
Microbiology lab.
|
S. No |
Action |
Responsibility |
|
4.1.1 |
All the glassware used in Micro lab should be
sterilized/dehydrogenated in oven per its SOP. Depyrogenation record is
maintained. |
Microbiologist |
|
S. No |
Action |
Responsibility |
|
4.2.1 |
Discard the used media and pure cultures by autoclave
and use of disinfectant as per SOP. |
Microbiologist |
|
S. No |
Action |
Responsibility |
|
4.3.1 |
Washing of glassware's and apparatus is performed after
discarding of media and samples with detergent and finally ruins with
distilled water. |
Microbiologist |
|
S. No |
Action |
Responsibility |
|
4.4.1 |
Place prepared and autoclaved media at below 8°c in refrigerator. Test samples
are also kept in the refrigerator. All pure cultures are also maintained and store
below 8°c.CSE is kept in
refrigerator after reconstitution. |
Microbiologist |
|
4.4.2 |
All Lysate tubes after reconstitution are kept at
freezer i.e below - 4°c or
below. Daily monitoring of both refrigerators QC-090 and -089
is performed on daily basis. |
|
S. No |
Action |
Responsibility |
|
4.5.1 |
Physical Balance is used to weight the media, testing
samples and other chemicals. Balance is calibrated daily prior use. It is
also calibrated on monthly basis.Quaterly calibration is performed by the
outside source (Scalemans) Records are maintained in respective log files. |
Microbiologist |
|
4.5.2 |
pH monitoring is performed of media prepared and
autoclaved. Buffers are prepared by monitoring their pH through pH meter. pH
meter is calibrated prior use by standards of pH 7.0 and 4.0.Record is
maintained in respective log files. |
|
S. No |
Action |
Responsibility |
|
4.6.1 |
All media used in Micro lab is must be sterilized by
Autoclave. All glassware and equipments that are highly heat sensitive are
sterilized by Autoclave .The glassware or other items that are not required
to be dry are also sterilized by Autoclave. |
Microbiologist |
|
4.6.2 |
Record of every Autoclave cycle is maintained with an
indicator. Validation of Autoclave sterilization cycles is performed on annual
basis. Operate the Autoclave as per SOP # QC |
|
S. No |
Action |
Responsibility |
|
4.7.1 |
Pure cultures are maintained by subculturing into fresh
slants at specified time and kept in
refrigerator below 8°c.These
pure cultures are used as positive controls in Microbial contamination, in
growth promoting test product suitability tests and Bioassays. |
Microbiologist |
|
4.7.2 |
List of Pure cultures and maintenance record of Pure
cultures is documented and maintained. |
Microbiologist |
4.8. Maintenance of Secondary standards:
|
S. No |
Action |
Responsibility |
|
4.8.1 |
All secondary standards of antibiotics used in bioassay
tests are kept in dessicator. They all are maintained in a list with C.A and expiry
dates. |
Microbiologist |
4.9. Cleaning and Maintenance of all
equipments:
|
S. No |
Action |
Responsibility |
|
4.9.1 |
Cleaning of all equipments is performed on daily basis
or as per requirement. Maintenance is also check at specified intervals. |
Microbiologist |
|
4.9.2 |
Cleaning and maintenance is check through monitoring at
specified intervals and monitoring record is maintained. |
4.10. Calibration of all equipments:
|
S. No |
Action |
Responsibility |
|
4.10.1 |
Calibration of all micro lab equipments is performed by
outside services company, Mecca Engineering and co. |
Microbiologist |
4.11. Maintenance of Sterility suit for sterility test:
|
S. No |
Action |
Responsibility |
|
4.11.1 |
Sterility suit is
maintained by one HEPA inside connected with HVAC. Positive pressure is
record by a magnahelic gauge. Temperature and humidity is also checked and
record by a hygrometer placed outside the. room |
Microbiologist |
|
4.11.2 |
Cleaning record is also
maintained of sterility suit. The areas is check by daily and weekly area
monitoring and record the results in specified Performa. |
|
|
4.11.3 |
Area of sterility suit
is also validated on annual basis. |
4.12. Bioassays:
|
S. No |
Action |
Responsibility |
|
4.12.1 |
These tests are performed to determine the potency of
antibiotics through microorganisms. These tests are performed as per
respective SOPs. |
Microbiologist |
4.13. Microbial contamination tests:
|
S. No |
Action |
Responsibility |
|
4.13.1 |
Microbial contamination tests are performed of all the
oral products including tablets, syrups, suspensions, dry syrups and
capsules.Microbail tests of all raw materials are also performed as per SOPs. |
Microbiologist |
4.17. Area Monitoring of Non sterile and
sterile areas:
|
S. No |
Action |
Responsibility |
|
4.17.1 |
Area monitoring is performed
in all sterile areas including Eye/Ear drops, Ampoule section and
Cephalosporin area on daily basis and Monthly basis as per SOP. The results
are delivered to all departments and allow production in these areas based on
the results. |
Microbiologist |
|
S. No |
Action |
Responsibility |
|
4.14.1 |
Animal house is maintained for Pyrogen testing. Healthy
rabbits are used for Pyrogen test. Proper housing is arranged for Rabbits and
cleaning is checked on daily basis. Their feeding record is also maintained
and All rabbits are physically checked at specified time and record is
maintained. |
Microbiologist |
|
4.9.2 |
Their injection history record is prepared with the
tests.Pyogen test is performed and reported as per the SOP and record in the
log book. |
|
S. No |
Action |
Responsibility |
|
4.15.1 |
This test is performed as per SOP of all Injectable manufactured
All Injectables are tested for sterility inside the maintained sterility suit. |
Microbiologist |
|
4.9.2 |
All Injectables should pass this test and products
released after 14 days of incubation. Daily monitoring of the test is
performed and recorded in the specified log book. |
|
S. No |
Action |
Responsibility |
|
4.16.1 |
To detect the presence of Endotoxin and its amount in
all Injectables products Endotoxin test is perform by Gel Clot method. |
Microbiologist |
|
4.16.2 |
LAL reagents are used to perform the test as per SOP
and record in the specified log books. The test is also perform for
manufacturing water of Ampoules and Eye/Ear drops and released for
manufacturing after passing the test. A separate record for manufacturing
water i.e. W.F.I is also maintained. |
|
S. No |
Action |
Responsibility |
|
4.17.1 |
Area monitoring is performed in all sterile areas
including Eye/Ear drops, Ampoule section and Cephalosporin area on daily basis
and on Monthly basis as per SOP. The results are delivered to all departments
and allow production in these areas based on the results. |
Microbiologist |
|
4.17.2 |
Area monitoring of all non sterile areas is performed
as per SOP on fortnightly basis. The results are documented and communicated
to concerned area Supervisors for advice and action. |
|
S. No |
Action |
Responsibility |
|
4.18.1 |
Microbial quality of water used in production of sterile
and nonsterile products is examined on daily basis. There is a weekly
examination of all water supply is carried out as per SOP. The results are
documented and action is taken as per the Alert and Action limits. |
Microbiologist |
|
S. No |
Action |
Responsibility |
|
4.19.1 |
Reports of all testing is transfer to QA deptt. for
release and approvals. All tests are performed as per the controlled SOPs and
documents are checked and revised periodically. |
Microbiologist |

0 Comments
If you have any doubt, please let me know.