STANDING OPERATION PROCEDURE (SOP) OF PERFORMANCE QUALIFICATION (PQ) OF INCUBATORS IN MICROBIOLOGY LAB
1.0. Purpose
The purpose of this
SOP is to provide the guideline for executing Performance Qualification (PQ) of
Incubators in order to demonstrate that the unit is capable of maintaining
overall temperature uniformity and stability through out its intended operating
requirement
2.0. Scope
This SOP
is applicable to all Incubators used in the Microbiology Laboratory.
3.0. Responsibilities
|
Action |
Responsibility |
|
Conducting
the qualification testing as per approved protocol, documentation of results
and the preparation of report. |
Microbiologist |
|
Calibration
of thermocouple and temperature recording device and assistance in carrying
out related tests |
Maintenance Dept. |
|
Compliance of SOP |
Laboratory
Incharge |
4.0. Policy Statement or Legal Requirements
N/A
General Requirements
4.1. Equipment
Instrumentation Calibration – All equipment instrumentation associated with the
incubators must be calibrated and remain in calibration during the
qualification test period. The unit should be placed on a preventative
maintenance (PM) program prior to initiating the performance qualification.
4.2. Test
Instrumentation Calibration – All test instrumentation used for the execution
of the PQ must be calibrated (i.e. Thermocouple & Temperature
Recording Device) and remain in calibration
during the qualification test period.
Performance Qualification of
incubators is carried out on annual basis in Microbiology Laboratory.
|
S. No |
Action |
Responsibility |
|
6.1.1 |
Place Five calibrated thermocouples at different locations within the chamber according to
map drawing in Annexure I - 2 |
Engineering
Dept. Microbiologist |
|
6.1.2 |
Plugged the main switch for power supply. |
Microbiologist |
|
6.1.3 |
Press
the ''GUIC ‘‘knob to ON the instrument indicated by Red light, Heater is also
“ON”. |
|
|
6.1.4 |
Adjust
the temperature by pressing “+” or “-”button to increase or decrease the set temperature indicated on
display monitor in the
whole number i.e. 22. 00C. |
|
|
6.1.5 |
To
“ON” fan inside press “Sigorta” Button indicated by orange light. |
|
|
6.1.6 |
Allow
the chamber to heat up until steady readings are observed. |
|
|
6.1.7 |
Temperature
data will be recorded for a period
of 12 hours with
temperature being logged at appropriate predefined time intervals. |
|
|
6.1.8 |
All
the temperature distribution reading will be obtained graphically/charted
during test length. |
|
|
6.1.9 |
After
completion of the temperature mapping, the final accumulated mean temperature at each location monitored during the test period will be calculated in
Annexure-1. |
|
|
6.1.10 |
Acceptance Criteria: The final accumulated mean temperature
of each location monitored during the dwell period should be within + 2.5
oC of set point for unit. |
|
S. No |
Action |
Responsibility |
|
6.2.1 |
Place Five calibrated thermocouples at different
locations within the chamber according to map drawing in Annexure -4 |
Engineering
Dept. Microbiologist |
|
6.2.2 |
Plugged the main switch to power supply. Make sure
the power supply by blinking the green light of “on” “off” button. |
Microbiologist |
|
6.2.3 |
Switch on the incubator by pressing the button mark
as “off”, directed from “o” to “I” |
|
|
6.2.4 |
Adjust the temperature by pressing the buttons mark
as “+” (for increase the digits) and “—“(for decrease the digits). |
|
|
6.2.5 |
Start stirrer (if required) by pressing the button
mark as “stirrer” directed from “0” to “I” |
|
|
6.2.6 |
Allow the chamber to heat up till steady readings
are observed. |
|
|
6.2.7 |
Temperature data will be recorded for a period of 12 hours with temperature being logged at
appropriate predefined time intervals. |
|
|
6.2.8 |
All the temperature distribution readings will be
obtained graphically/charted during test length. |
|
|
6.2.9 |
After completion of the temperature mapping, the
final accumulated mean temperature at each location monitored during the test
period will be calculated in Annexure – |
|
|
6.2.10 |
Acceptance Criteria:
The final accumulated mean temperature of each location monitored during the
dwell period should be within + 2.5oC of set point for
unit. |
|
S. No |
Action |
Responsibility |
|
6.3.1 |
Place Five calibrated
thermocouples at different locations within the chamber according to map
drawing in Annexure. |
Engineering
Dept. Microbiologist |
|
6.3.2 |
Plugged
the main switch to power supply. |
Microbiologist |
|
6.3.3 |
Turn “ON” the '' ON ''
“OFF” knob, directed from “o” to “I” |
|
|
6.3.4 |
Adjust the temperature by
rotate the temperature rotatory knob as required temperature. |
|
|
6.3.5 |
Allow the chamber to heat
up until steady readings are observed. The chamber temperature has stabilized
when there is uniform cycling of the control green pilot light. |
|
|
6.3.6 |
Temperature data will be
recorded for a period of
12 hours with temperature being logged at appropriate predefined time
intervals. |
|
|
6.3.7 |
All the temperature
distribution reading will be obtained graphically/charted during test length.
|
|
|
6.3.8 |
After completion of the
temperature mapping, the final accumulated mean temperature at each location
monitored during the test period will be calculated in Annexure- |
|
|
6.3.9 |
Acceptance Criteria:
The final accumulated mean temperature of each location monitored during the
dwell period should be within + 2.5oC of set point for
unit. |
|
S. No |
Action |
Responsibility |
|
6.4.1 |
Place Five calibrated thermocouples at different locations within the
chamber according to map drawing in Annexure -8. |
Engineering
Dept. Microbiologist |
|
6.4.2 |
Switch on the power supply by pressing the
”ON”, OFF” knob of main control panel. |
Microbiologist |
|
6.4.3 |
Press
the power switch to “ON” the incubator. |
|
|
6.4.4 |
Adjust the set temperature (required temperature) by pressing “UP” and
“DOWN” keys of main control panel. |
|
|
6.4.5 |
Check the “keep warm” and “Disinfecting” status by pressing “KEEP
WARM” and “DISINFECTING” knob and confirm by blinking the “ keep warm light” (yellow
light) and “disinfecting light” (red light),present at the base of incubator. |
|
|
6.4.6 |
Temperature data will be recorded for a period of 12 hours with
temperature being logged at appropriate predefined time intervals. |
|
|
6.4.7 |
All the temperature distribution reading will be obtained
graphically/charted during test length. |
|
|
6.4.8 |
After completion of the temperature mapping, the final accumulated
mean temperature at each location monitored during the test period will be
calculated in Annexure- |
|
|
6.4.9 |
Acceptance Criteria:
The final accumulated mean temperature of each location monitored during the
dwell period should be within + 1.0 oC of set point for
unit. |
7.0. Definitions
|
Term/word/statement |
Meaning/Description |
|
Performance
qualification |
Identification of equipment attributes related to the
performance of a particular function or functions and allocation of certain
limits or restriction to those attributes. |
|
S. No |
Description |
|
1. |
N/A |
|
S. No |
Description of Record |
|
1. |
N/A |
|
Document No |
Description |
|
|
N/A |
|
S. No |
Description of Annexure |
|
1. |
Temperature Recording during P.Q of Incubator Name (QC-00) |
|
2. |
Thermocopule Position Map of Incubator name (QC- 000) |
|
3. |
Temperature
Recording during P.Q of Incubator name (QC- 000) |
|
4. |
Thermocopule Position
Map of Incubator name (QC- 000) |
|
5. |
Temperature
Recording during P.Q of Incubator name (QC- 000) |
|
6. |
Thermocopule Position
Map of Incubator name (QC- 000) |
|
7. |
Temperature
Recording during P.Q of Incubator name (QC - 000) |
|
8. |
Thermocopule Position Map of Incubator name (QC - 000) |
Distribution List:
|
1. Head Quality Assurance |
2. Incharge Microbiology |